Under what circumstances should you use Doppler flowmeters? - Part 1

May 2001 Flow Measurement & Control

Clamp-on ultrasonic Doppler flowmeters are used in a wide variety of industrial, chemical, wastewater and food processing applications. Understanding how these nonintrusive flowmeters work will help you to understand when to use them.

Clamp-on ultrasonic Doppler flowmeters flow in full pipes containing slurries, sludges, emulsions, dispersions and pulps - any liquid containing suspended particles or droplets of a material with a sonic impedance that is different than the sonic impedance of the carrier liquid.

To assess useful applications for Doppler flowmeters, it is necessary to understand the basic operating principle of the meter and some understanding of relevant ultrasound scattering theory.

Operating principles

Most Doppler flowmeters use separate transmit and receive transducers clamped onto opposite sides of a pipe exterior. An ultrasonic beam is transmitted through the pipe wall into the flowing fluid at an oblique angle to the flow. The intercept of the transmit and receiver ultrasound beams is directed at the centre of the flow. The receiver collects scattered ultrasound from a core group of flowing particles travelling down the central axis of the pipe. This received ultrasound is shifted in frequency from the transmitted ultrasound in accordance with the general Doppler equation:

Equation 1
Equation 1

Figure 1 depicts a simplified, two-dimensional diagram of the ultrasound beams injected into the flow stream with the centre of the beam at an angle. To simplify this explanation, let us only consider the velocities along the centreline of flow.

Figure 1
Figure 1

The received Doppler shifted frequencies mix with the transmit frequency, resulting in a group of beat frequencies. These time domain, sinusoidal beat signals are digitised and converted to the frequency domain via a fast Fourier transform (FFT). The frequency spectrum generated by the FFT is scanned to find the dominant peak frequency. Figure 2 illustrates a typical frequency distribution from a Doppler flowmeter. The point of highest intensity, the frequency spectrum peak, indicates the centre flow velocity.

Figure 2
Figure 2

The centre flow velocity is then converted to average flow velocity using flow profile correction factors based on Reynolds number and pipe roughness. Profile corrections allow the meter to maintain accuracy when transferred from one pipe size and material to another. The average flow is determined by multiplying the internal cross-sectional area of the pipe by the average fluid velocity.

What constitutes a reasonable Doppler application?

Sonic scatterers

Doppler flowmeters are not generally promoted for use on clean fluids. While powerful Doppler flowmeters can read the minute particles in municipal drinking water on small pipes under laboratory conditions, Doppler flowmeters cannot read the flow of highly distilled water.

In real applications, strong Doppler shifted signals are required for reliable flow measurement. The amplitude of the Doppler shifted signals from any application is largely related to the suitability of the sonic scatterers in the flow for producing detectable Doppler signals.

Identifying suitable sonic scatterers for Doppler flowmeters is a complex combination of four basic criteria. How well the criteria are met determines the reliability and accuracy of a Dopper flow application. The four criteria are:

1. The scattering material must have a sonic impedance different from the fluid.

2. There must be some particles large enough to cause longitudinal scattering.

3. For a given pipe size, the longitudinal scattering must have sufficient energy to overcome the Rayleigh (energy wasting) scattering caused by smaller particles.

4. The scattering material must travel at the same velocity as the fluid for good accuracy.

The sonic impedance of a material is the product of the density and the sound speed. An ideal Doppler scatterer has the same density as the fluid, but a different sound speed. Polyethylene particles, for example, can be used in water for testing and calibration purposes. The polyethylene has a density nearly identical to water, and a sound speed one-third higher, so the overall impedance is one-third higher. While some impedance difference is required, the dominant characteristic defining useful scatterers is particle diameter. The 500 µm diameter of the polyethylene particles creates very strong longitudinal scattering, with a concentration of only 0,025% by volume. The similar density of the materials keeps the particles in a suspended slurry.

To present useful guidelines, the following discussion will use water as the liquid carrier and 500 kHz as the transmission frequency (a common Doppler transmission frequency).

Particles less than 50 µm (0,002") fall in the Rayleigh regime, and only waste ultrasound energy by thermoelastic and viscoinertial scattering. High concentrations of particles in this range are extremely difficult to measure by any ultrasound method.

Particles greater than 0,002" and less than one-eighth of an inch, create longitudinal, thermoelastic and viscoinertial scattering. The longitudinal energy increases with the particle diameter to the fourth power (d4) in this regime. In a dispersed particle distribution, a few large particles contribute the scattering needed for reliable Doppler signals while the smaller particles waste ultrasound energy. Because the ultrasound energy must travel through this highly attenuating media, pipe diameter becomes an important variable. A particular particle distribution generating sufficient Doppler signals in a small pipe may not produce sufficient signals in a larger pipe.

Particles of diameter greater than one-eighth of an inch cause high energy geometric scattering. Clamp-on Doppler flowmeters easily measure flows carrying small concentrations of particles in this size range.

In real applications, there is often a distribution of particle sizes. Experience has shown that if the volume of particles greater than 100 µm are 25% of the overall particle volume, the system is adequate for a Dopper flowmeter on moderate size pipes (ie 6"). For larger pipes, the distributions may require 50% or higher.

An example is a slurry of water and calcium sulphate particles used for industrial stack scrubbing. The concentration is 4% by volume, pumped through a 6" fibreglass pipe. Most of the particles by count are very small and dissipate a great deal of the ultrasonic energy transferred into the slurry. However, 35% of the particles by particle volume are 100 µm diameter and greater, as depicted in Figure 3.

Figure 3
Figure 3

The percent of large particles by volume provides sufficient longitudinal scattering to overcome the dissipation effects of the smaller particles. While this application had low signal strength and a wide frequency spectrum, a reasonable and stable flow measurement was achieved.

Doppler flowmeters measure particle velocity and are dependent on scattering material travelling at the same velocity as the fluid for accuracy. How will particle size and density affect accuracy? For pulps and sludges, the sonic scatterers tend to be fibrous material of density less than or equal to water in high concentration. These fibrous materials create large drag forces, causing them to follow the flow streams of the fluid at the same velocity despite flow disruptions.

Titan Technology

(011) 888 3009

[email protected]





Share this article:
Share via emailShare via LinkedInPrint this page

Further reading:

Radiometric density measurements
Mecosa Flow Measurement & Control
Radiometric density measurements are used wherever conventional measurement techniques fail. Successful applications can be found in a wide variety of industries, especially where harsh process environments and difficult measuring conditions prevail.

Read more...
Battery-powered magmeter
Flow Measurement & Control
The RWEM/D battery-powered electromagnetic flow meter from Royce Water Systems is ideal for water and wastewater systems located at remote sites.

Read more...
Precise process and maximum savings with VEGABAR
VEGA Controls SA Flow Measurement & Control
Accurate measurementsof differential pressure is crucial for safety, efficiency and optimising operational processes. For over 40 years, VEGA has supplied industry-leading pressure measurement solutions, establishing itself as a trusted expert with a proven track record in sensor technology.

Read more...
Innovative solutions for low flow dosing with high precision
Mecosa Flow Measurement & Control
n the food, pharma, and chemical manufacturing industries, precision and efficiency are of utmost importance. Bronkhorst High-Tech has developed a flexible solution that meets today’s requirements for highly accurate dosing, whether continuous, in ratio, or in batches.

Read more...
Battery-powered magmeter
Flow Measurement & Control
The RWEM/D battery-powered electromagnetic flow meter from Royce Water Systems is ideal for water and wastewater systems located at remote sites.

Read more...
Smart technology for municipal services
VEGA Controls SA Flow Measurement & Control
A new wave of smart technologies has emerged from VEGA Instruments, promising a revolutionary impact on the efficiency and dependability of municipal services.

Read more...
Sustainable journey with hydrogen
Mecosa Flow Measurement & Control
Bronkhorst High-Tech, the Dutch leader in flow measurement and control, has been focusing on hydrogen applications in the renewable energy market for some time, and for good reason.

Read more...
Optimise sludge treatment and reduce operational costs
Endress+Hauser South Africa Flow Measurement & Control
The Endress+Hauser inline measuring devices, the Proline Teqwave MW 300 and the Proline Teqwave MW 500, determine the total solids content of wastewater directly through microwave transmission.

Read more...
No more blind spots in flow measurement of liquid foods
ifm - South Africa Flow Measurement & Control
The SM Foodmag magnetic-inductive flow sensor takes the flow measurement of liquid and creamy foods to a new level.

Read more...
Plant signal conditioning
Omniflex Remote Monitoring Specialists Flow Measurement & Control
Often, plant managers have a deep understanding of all the key operational challenges facing their facility, and have a broad knowledge of the instrumentation that can help solve them. However, they might not always be aware of the latest innovations that can make their lives even easier.

Read more...